Radia Diagnostic Modules & Tests

Click to download a PDF of this information.

Module and Test Radia Outputs

Catphan, kV & Cone Beam Phantoms

Complete Automatic scoring of all phantom parameters.

Catphan 500/600

Measures and plots material values and CT number linearity. Measures: slice width, ramp alignment, rod distances, low contrast detectability, measures uniformity, integral non-uniformity, plots integral non-uniformity profile, contrast-to-noise, scores low contrast objects, measures and plots MTF.

ACR CT Analysis (Follows 10/2011 guidelines)

BB location measurements, slice thickness, CT material accuracy, CT material uniformity, contrast-to-noise ratio, contrast statistics, signal-to-noise, uniformity measurements

Additional analysis: MTF (SD Method), modulation measurement and plot. Detection of low contrast rods.

Measure for all four protocols (Adult Abdomen, Adult Head, Pediatric Abdomen, Pediatric Head).

CT Material analysis may be done on any Module 1 slice!

ACR MRI Analysis

Localizer line length, geometric accuracy (slice 1 and slice 5), resolution, slice thickness, slice position accuracy (slice 1 & 11), Percent integral uniformity (with full statistics), Percent signal ghosting (with full statistics), Low contrast objects detected with contrast-to-noise reported for each object.

ACR Small MRI Analysis (Extremity & Breast) 

Localizer line length, geometric accuracy (slice 1 and slice 3), resolution, slice thickness, slice position accuracy, percent integral uniformity (with full statistics), percent signal ghosting (with full statistics), low contrast objects detected with contrast-to-noise reported for each object.

DISC Plus Phantom, DR/CR/Fluoro kV Imaging
IBA Primus L Phantom (30cmx30cm test area! – LARGE AREA testing)

Mean, standard deviation and contrast ratio for each of the 17 steps. Contrast-to-noise ratio for each of the 8 low contrast objects. kV measurement with Ytterbium dot, uniformity and noise of center area, modulation reported and plotted for the resolution pattern, geometric distortion measurement.

ACR FFDM (available when phantom is released)

Complete automatic scoring the new ACR FFDM to full requirements. Scores fibers, specs and masses to half a point. Also scores contrast to noise, signal to noise, and insert size.

DICOM image viewer and DICOM tag viewer

Several methods to locate and import your images

DICOM directory browser, open directory browser, DICOM file filter for organizing your images

RunQueueC

Batch processing of images

Statistical Trend database

All measurements can be tracked and plotted over time in the statistical trend database. Any number of databases can be used. All Statistical Process Control parameters measured and plotted.

Monitor Image Quality

SMPTE pattern and TG18 images

Manual Measurements

Distance, angle, profile, histograms, round and square ROI. ASCII output of profiles.

Output formats

Print, PDF, Excel (with custom templates), XML

Image Transfer

Built in image FTP function