| Module and Test | Radia Outputs | |
|
Catphan/OBI, kV/MV & Cone Beam CT Phantoms |
Complete Automatic scoring of all phantom parameters. |
|
|
Varian 504 Catphan |
Measures and plots material values and CT number linearity. Measures: slice width, ramp alignment, rod distances, low contrast detectability, measures uniformity, integral non-uniformity, plots integral non-uniformity profile, contrast-to-noise, scores low contrast objects, measures and plots MTF. |
|
|
Elekta 503 Catphan |
Measures and plots material values and CT number linearity. Measures: slice width, ramp alignment, rod distances, low contrast detectability, measures uniformity, integral non-uniformity, plots integral non-uniformity profile, measures and plots MTF, uniformity mean difference, low contrast visibility. |
|
|
Siemens Image Quality Phantom |
Geometrical measurements, central area mean and standard deviation, uniformity measurements, measures and plots MTF, low contrast target scoring. |
|
|
Catphan 500/600 |
Measures and plots material values and CT number linearity. Measures: slice width, ramp alignment, rod distances, low contrast detectability, measures uniformity, integral non-uniformity, plots integral non-uniformity profile, contrast-to-noise, scores low contrast objects, measures and plots MTF. |
|
|
Electron Density/Tissue Characterization Phantoms |
||
|
CIRS 062 + 062A (cone beam) |
Measures and plots 17 density areas. |
|
|
Gammex 467 |
Measures and plots 17 density areas. Geometric measurement in X and Y direction. |
|
|
EPID QC |
||
|
PTW EPID QC Phantom (26cmx26cm test area at isocenter! – LARGE AREA testing) |
Mean, standard deviation, and signal to noise for each of the 10 copper steps. Mean value for each of the four objects in the six brass steps, horizontal and vertical spacing for geometric accuracy, MTF reported and plotted for resolution pattern, contrast is reported for each of the 27 contrast objects. |
|
|
DR/CR/Fluoro kV Imaging |
||
|
DISC Plus Phantom, IBA Primus L Phantom (30cmx30cm test area! – LARGE AREA testing) |
Mean, standard deviation and contrast ratio for each of the 17 steps. Contrast-to-noise ratio for each of the 8 low contrast objects. kV measurement with Ytterbium dot, uniformity and noise of center area, modulation reported and plotted for the resolution pattern, geometric distortion measurement. |
|
|
DICOM image viewer and DICOM tag viewer |
||
|
Statistical Trend database |
All measurements can be tracked and plotted over time in the statistical trend database. Any number of databases can be used. All Statistical Process Control parameters measured and plotted. |
|
|
Monitor Image Quality |
SMPTE pattern and TG18 images |
|
|
Manual Measurements |
Distance, angle, profile, histograms, round and square ROI. ASCII output of profiles. |
|
|
Output formats |
Print, PDF, Excel (with custom templates), XML |
|
|
Image Transfer |
Built in image FTP function |
|